@@ -10,6 +10,11 @@ diffraction, and computes the transmission `T` and normalised beam deflection
compared against the simulation, and the result exported as data, a log, and
publication-sized figures.
Two measurement modes are supported. **Beam deflection** fits both signals.
**Pump probe** is for measurements with the probe centred on the pump, where
there is no deflection: it fits the transmission alone, and needs only delay
and transmission data.
![BD Fresnel fitting a measured scan](docs/screenshot.png)
## Features
@@ -23,6 +28,9 @@ publication-sized figures.
- Excited-state absorption via a five-level population model (S₀, S₁, S₂, T₁,
T₂) solved through the sample depth for a range of radial irradiances.
- Side-by-side transmission and deflection plots with imported data overlaid.
- Beam deflection and pump probe measurement modes. Pump probe reads a
two-column delay/transmission scan, fixes the probe on the pump axis, and
offers only the mechanisms that can change the transmission.
- Optional diagnostic figures showing each mechanism's response function before
and after convolution, plus the ESA populations and beam transmissions.
- Export of the data/fit comparison, a full parameter log, and figures at
@@ -57,29 +65,36 @@ right: **Fitting**, **Diagnostics**, and **Parameter log**.
A typical session:
1. Set the **Data columns** to match your file, and the delimiter.
2. Press **Import…** and choose a scan. The plots show the measured data.
3. Fill in the beam, material and geometry parameters, and configure one or
2. Choose the **Mode**: Beam deflection or Pump probe.
3. Press **Import…** and choose a scan. The plots show the measured data.
4. Fill in the beam, material and geometry parameters, and configure one or
more mechanisms with the **Params…** buttons.
4. Press **Fit** to run the simulation and overlay it on the data.
5. Press **Export…** to write the results.
5. Press **Fit** to run the simulation and overlay it on the data.
6. Press **Export…** to write the results.
**Fit** stays disabled until a file is imported, and **Export** until a fit has
completed.
Changing the mode, a column number or the delimiter re-reads the imported file.
Any fit is discarded, since it belonged to the previous settings. If the file
cannot be read with the new settings, its data is cleared rather than left on
screen, and it is read again once the settings allow.
Parameters are saved to `last_settings.json` beside the scripts when they
change, and restored at the next launch. The data file itself is not restored.
## Input data format
A delimited text file with one header line, then one row per delay point. Three
columns are required; which ones is set in the **Data columns** box, so extra
columns can be present and ignored.
A delimited text file with one header line, then one row per delay point. Beam
deflection needs three columns and pump probe two; which ones is set in the
**Data columns** box, so extra columns can be present and ignored. Each
quantity must come from a different column.
| Quantity | Units | Default column |
| --- | --- | --- |
| `τ_d` | fs | 1 |
| `ΔE/E` | fractional | 2 |
| `T` | fractional (1.0 = no change) | 3 |
| Quantity | Units | Default column | Needed in |
| --- | --- | --- | --- |
| `τ_d` | fs | 1 | both modes |
| `ΔE/E` | fractional | 2 | Beam deflection only |
| `T` | fractional (1.0 = no change) | 3 | both modes |
The delimiter field takes a single character, `\t` for a tab, or is left blank
to split on any run of whitespace.
@@ -95,6 +110,13 @@ to split on any run of whitespace.
| `DeltaEoE column` | `2` | Deflection data column |
| `Delimiter` | `,` | Data column delimiter character |
### Mode
| Option | Effect |
|-------------------|-------------------------------------------------|
| `Beam deflection` | Transmission and deflection, offset probe |
| `Pump probe` | Transmission only, centered probe |
### Simulation parameters
| Variable | Default | Column contents |
@@ -141,7 +163,7 @@ to split on any run of whitespace.
| `n` | `1.4565` | Linear index |
| `dndl` | `-0.027158` | First order dispersion (um^-1) |
### System geometry
### Measurement parameters
| Variable | Default | Column contents |
|----------------|-------------|-----------------------------------|
@@ -167,24 +189,31 @@ to split on any run of whitespace.
`NLR` denotes a purely refractive response, `NLA` a purely absorptive one, and
`both` a mechanism carrying each.
Pump probe offers only the `NLA` and `both` mechanisms: with the probe centred,
a `both` mechanism contributes its absorptive part, and an `NLR` mechanism
cannot change the transmission at all. Switching to Pump probe sets aside any
`NLR` mechanism already chosen; switching back restores it, provided its slot
is still empty.
## Export output
Exporting to `<name>_fit.txt` also writes, in the same directory:
| File | Contents |
| --- | --- |
| `<name>_fit.txt` | delay, data and fit for `T`, `ΔE/E` and `Δn` |
| `<name>_fit.txt` | delay, data and fit for `T`, `ΔE/E` and `Δn`; in Pump probe, delay and `T` only |
| `<name>_fit_log.txt` | every parameter, plus a manifest of the files written |
| `<name>_fit_combined.jpg` / `.svg` | transmission and deflection, 6.5 in at 300 dpi |
| `<name>_fit_combined.jpg` / `.svg` | transmission and deflection, 6.5 in at 300 dpi (Beam deflection) |
| `<name>_fit_transmission.jpg` / `.svg` | transmission only, 3.25 in at 300 dpi (Pump probe) |
| `<name>_fit_raw_response.jpg` | each mechanism's `R[t]` and `A[t]` |
| `<name>_fit_convolved_response.jpg` | the same after convolution, plus the total |
The last two appear only when *Show diagnostic plots during Fit* is ticked.
The log records enough to reproduce a run: data columns and delimiter, the
simulation grid, every beam, material and geometry value, each mechanism with
its parameters, and the derived quantities (`n_g`, `T_lin`, `z_0`, `I_0`,
`rho`).
mode, the simulation grid, every beam, material and geometry value, each
mechanism with its parameters, and the derived quantities (`n_g`, `T_lin`,
`z_0`, `I_0`, `rho`).
## Files
@@ -225,4 +254,4 @@ You should have received a copy of the GNU General Public License along with
this program. If not, see <https://www.gnu.org/licenses/>.
NumPy, SciPy, Matplotlib and mpmath are all BSD-licensed, so distributing this program
under the GPL alongside them poses no conflict.
under the GPL alongside them poses no conflict.