# BD Fresnel Beam-deflection pump–probe simulator and fitting tool. A Python port of the MATLAB `BD_Fresnel_GUI` suite. The program models a weak probe beam crossing a sample that has been excited by a strong pump, propagates the perturbed probe field to a detector by Fresnel diffraction, and computes the transmission `T` and normalised beam deflection `ΔE/E` as a function of pump–probe delay. Measured scans can be imported and compared against the simulation, and the result exported as data, a log, and publication-sized figures. Two measurement modes are supported. **Beam deflection** fits both signals. **Pump probe** is for measurements with the probe centred on the pump, where there is no deflection: it fits the transmission alone, and needs only delay and transmission data. The program also runs without a measurement. In **Simulation** mode nothing is imported: the signal is computed from the parameters alone, over a delay range entered by hand, and only the calculated curves are plotted and exported. ![BD Fresnel fitting a measured scan](docs/screenshot.png) ## Features - Seven mechanism slots, each independently configurable, summed into the total nonlinear response. - Instantaneous (electronic) and non-instantaneous responses, the latter convolved with the pulse envelope. - Group-velocity mismatch between pump and probe handled through the walk-off parameter `rho`. - Excited-state absorption via a five-level population model (S₀, S₁, S₂, T₁, T₂) solved through the sample depth for a range of radial irradiances. - Side-by-side transmission and deflection plots, with imported data overlaid when there is any. - Beam deflection and pump probe measurement modes. Pump probe reads a two-column delay/transmission scan, fixes the probe on the pump axis, and offers only the mechanisms that can change the transmission. - Fitting and simulation run modes. Simulation computes the signal with no data imported, over a delay range entered in the panel. - Optional diagnostic figures showing each mechanism's response function before and after convolution, plus the ESA populations and beam transmissions. - Export of the data/fit comparison, a full parameter log, and figures at 6.5 in / 300 dpi. ## Requirements - Python 3.10 or newer (developed against 3.12) - `numpy`, `scipy`, `matplotlib`, `mpmath` - Tkinter, which ships with most Python builds. On Debian/Ubuntu it is a separate package: ```sh sudo apt install python3-tk ``` Install the Python dependencies with: ```sh pip install -r requirements.txt ``` ## Running ```sh python bd_fresnel_gui.py ``` The window opens with the parameter panel on the left and three tabs on the right: **Fitting**, **Diagnostics**, and **Parameter log**. Fitting a measured scan: 1. Under **Operation mode**, choose **Beam deflection** or **Pump probe**, and leave the second toggle on **Fitting**. 2. Set the **Data columns** to match your file, and the delimiter. 3. Press **Import…** and choose a scan. The plots show the measured data. 4. Fill in the beam, material and geometry parameters, and configure one or more mechanisms with the **Params…** buttons. 5. Press **Fit** to run the simulation and overlay it on the data. 6. Press **Export…** to write the results. Simulating without a measurement: 1. Under **Operation mode**, choose **Simulation**. **Import…** and the data column fields are disabled, and **Fit** becomes **Simulate**. 2. Set `T_d_min` and `T_d_max` in **Simulation parameters** to the delay range to compute over. In Fitting mode these come from the imported file, so they are editable only here. 3. Fill in the parameters and mechanisms as above. 4. Press **Simulate**, then **Export…**. **Fit** stays disabled until a file is imported; **Simulate** needs only valid parameters. **Export** stays disabled until a fit or simulation has completed. Changing the mode, a column number or the delimiter re-reads the imported file. Any fit is discarded, since it belonged to the previous settings. If the file cannot be read with the new settings, its data is cleared rather than left on screen, and it is read again once the settings allow. Parameters are saved to `last_settings.json` beside the scripts when they change, and restored at the next launch. The data file itself is not restored. ## Input data format A delimited text file with one header line, then one row per delay point. Beam deflection needs three columns and pump probe two; which ones is set in the **Data columns** box, so extra columns can be present and ignored. Each quantity must come from a different column. | Quantity | Units | Default column | Needed in | | --- | --- | --- | --- | | `τ_d` | fs | 1 | both modes | | `ΔE/E` | fractional | 2 | Beam deflection only | | `T` | fractional (1.0 = no change) | 3 | both modes | The delimiter field takes a single character, `\t` for a tab, or is left blank to split on any run of whitespace. ## Parameters ### Data columns (1-indexed) | Variable | Default | Column contents | |-------------------|---------|---------------------------------------| | `tau_d column` | `1` | Delay column (fs) | | `T column` | `3` | Transmission data column | | `DeltaEoE column` | `2` | Deflection data column | | `Delimiter` | `,` | Data column delimiter character | ### Operation mode | Option | Effect | |-------------------|-------------------------------------------------| | `Beam deflection` | Transmission and deflection, offset probe | | `Pump probe` | Transmission only, centered probe | | `Fitting` | Compare the calculation against an imported scan | | `Simulation` | Calculate without a scan, over an entered range | ### Simulation parameters | Variable | Default | Column contents | |----------------|---------|---------------------------------------| | `X_num` | `50` | Number of spatial points | | `X_max` | `5.0` | Range of spatial sampling (w_0) | | `T_num` | `100` | Number of temporal points | | `T_max` | `5.0` | Range of temporal sampling (tau_p) | | `T_d_num` | `150` | Number of delay points over range | | `T_d_min` | `-5.0` | Delay range start (tau_p), Simulation only | | `T_d_max` | `5.0` | Delay range end (tau_p), Simulation only | ### Excitation beam parameters | Variable | Default | Column contents | |----------------|---------|---------------------------------------| | `E_p` | `1.36` | Excitation pulse energy (uJ) | | `tau_p_FWHM` | `55.0` | Excitation pulse duration (fs) | | `w_0p` | `246.0` | Excitation beam waist (um) | | `lambda_p` | `800.0` | Excitation pulse wavelength (nm) | ### Probe beam parameters | Variable | Default | Column contents | |----------------|---------|---------------------------------------| | `tau_FWHM` | `60.0` | Probe pulse duration (fs) | | `w_0` | `43.0` | Probe beam waist (um) | | `lambda` | `650.0` | Probe pulse wavelength (nm) | ### Material parameters at excitation wavelength | Variable | Default | Column contents | |----------------|-------------|-----------------------------------| | `R_p` | `0.0` | Reflectivity | | `alpha_p` | `0.0` | Linear absorption (cm^-1) | | `n_p` | `1.4533` | Linear index | | `dndl_p` | `-0.017284` | First order dispersion (um^-1) | | `alpha_2d` | `0.0` | 2 photon absorption (e-11 m/W) | ### Material parameters at probe wavelength | Variable | Default | Column contents | |----------------|-------------|-----------------------------------| | `R` | `0.0` | Reflectivity | | `alpha` | `0.0` | Linear absorption (cm^-1) | | `n` | `1.4565` | Linear index | | `dndl` | `-0.027158` | First order dispersion (um^-1) | ### Measurement parameters | Variable | Default | Column contents | |----------------|-------------|-----------------------------------| | `l` | `1.0` | Sample thickness (mm) | | `d` | `15.0` | Distance to detector (cm) | | `shift` | `0.0` | Temporal shift (fs) | | `x_0` | `0.5` | Displacement of probe (w_0p) | ## Mechanisms | Mechanism | Parameters | | --- | --- | | Electronic (both) | 2 | | Collisional (NLR) | 3 | | Librational (NLR) | 4 | | Reorientational (NLR) | 3 | | Vibrational (NLR) | 3 | | ESA – 3 level + triplet (both) | 10 | | Exponential rise/decay (NLA) | 3 | | Exponential rise/decay (NLR) | 3 | | Exponential rise/decay (both) | 4 | `NLR` denotes a purely refractive response, `NLA` a purely absorptive one, and `both` a mechanism carrying each. Pump probe offers only the `NLA` and `both` mechanisms: with the probe centred, a `both` mechanism contributes its absorptive part, and an `NLR` mechanism cannot change the transmission at all. Switching to Pump probe sets aside any `NLR` mechanism already chosen; switching back restores it, provided its slot is still empty. ## Export output Exporting to `_fit.txt` also writes, in the same directory: | File | Contents | | --- | --- | | `_fit.txt` | delay, data and fit for `T`, `ΔE/E` and `Δn`; in Pump probe, delay and `T` only. In Simulation, the calculated signals alone | | `_fit_log.txt` | every parameter, plus a manifest of the files written | | `_fit_combined.jpg` / `.svg` | transmission and deflection, 6.5 in at 300 dpi (Beam deflection) | | `_fit_transmission.jpg` / `.svg` | transmission only, 3.25 in at 300 dpi (Pump probe) | | `_fit_raw_response.jpg` | each mechanism's `R[t]` and `A[t]` | | `_fit_convolved_response.jpg` | the same after convolution, plus the total | The last two appear only when *Show diagnostic plots during Fit* is ticked. In Simulation mode the export is named after the file you choose rather than an imported scan, and the table holds the calculated signals on their own delay grid, with no data columns to pair them with. The log records enough to reproduce a run: both modes, the data columns and delimiter, the simulation grid including the delay range, every beam, material and geometry value, each mechanism with its parameters, and the derived quantities (`n_g`, `T_lin`, `z_0`, `I_0`, `rho`). In Simulation mode the import path and data column sections are omitted, having nothing to describe. ## Files | File | Role | | --- | --- | | `bd_fresnel_core.py` | physics: field propagation, mechanism responses, ESA solver, parameter dataclass, settings I/O. No display dependencies. | | `bd_fresnel_gui.py` | Tkinter interface, plotting, import/export, logging. Entry point. | | `set_params_dialog.py` | modal dialog for entering one mechanism's parameters. | ## References Please cite the following when using results from this program. 1. Reichert, M., et al. (2014). "Temporal, spectral, and polarization dependence of the nonlinear optical response of carbon disulfide." *Optica* **1**(6): 436. 2. Reichert, M., et al. (2014). "Temporal, spectral, and polarization dependence of the nonlinear optical response of carbon disulfide: supplementary material." *Optica* **1**(6): 436. 3. Ferdinandus, M. R., et al. (2013). "Beam deflection measurement of time and polarization resolved ultrafast nonlinear refraction." *Opt. Lett.* **38**(18): 3518–3521. ## License Copyright (C) 2019-2026 Manuel R. Ferdinandus This program is free software: you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation, either version 3 of the License, or (at your option) any later version. This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the GNU General Public License for more details. You should have received a copy of the GNU General Public License along with this program. If not, see . NumPy, SciPy, Matplotlib and mpmath are all BSD-licensed, so distributing this program under the GPL alongside them poses no conflict.