4 Commits
Author SHA1 Message Date
manuel 29b827e307 Update README.md 2026-09-22 01:09:34 +00:00
manuel 17179f32c9 Update README.md 2026-09-21 10:56:15 +00:00
manuel 1bf80b2f9d Update README.md 2026-09-17 16:35:36 +00:00
manuel 79bfcd8084 Merge pull request 'Add readme' (#1) from add_readme into main
Reviewed-on: #1
2026-09-12 19:57:28 +00:00
+75 -17
View File
@@ -141,33 +141,91 @@ The program is split so the physics can be used without the interface:
## Parameters ## Parameters
<details> ### Data columns (1-indexed)
<summary>Full parameter list</summary>
**Data columns** — `z_col`, `CA_col`, `OA_col`, `REF_col`, `FREQ_col` (1-indexed) | Variable | Default | Column contents |
|------------|---------|---------------------------------------|
| `z_col` | `0` | Sample position *z* (mm) |
| `CA_col` | `1` | Closed-aperture detector signal (V) |
| `FREQ_col` | `3` | Frequency / shot-index channel |
| `OA_col` | `4` | Open-aperture detector signal (V) |
| `REF_col` | `7` | Reference detector signal (V) |
| `FREQ_col` | `9` | Lock-in frequency column (V) |
**Processing flags** — `use_ref`, `use_freq`, `flip_z`, `do_norm_OA`, ### Processing flags
`do_norm_CA`, and per-channel sign flips for the raw and displayed traces
**Nonlinear coefficients** — `n_2` (m²/W), `n_3` (m⁴/W²), `n_4` (m⁶/W³), | Parameter | Meaning |
`alpha_2` (m/W), `alpha_3` (m³/W²), `alpha_4` (m⁵/W³) |--------------------------------------|-------------------------------------------------------------|
| `use_ref` | Apply the reference-channel correction |
| `use_freq` | Appliy the frequency-channel correction |
| `flip_z` | `1` reverses the sign of the z axis |
| `do_norm_OA`, `do_norm_CA` | Fit and divide out the linear wing baseline |
| `flip_OA`, `flip_CA`, `flip_REF` | Invert a raw channel about its median (`0` or `1`) |
| `flip_OA_disp`, `flip_CA_disp` | Invert the normalized traces about 1 for display/fitting |
| `show_CA` | Plot CA in the right panel; if `False`, plot CA/OA instead |
**z-position offsets** — `Delta_z_OA`, `Delta_z_CA`, `Delta_z_OACA` (mm), to ### Nonlinear coefficients
align each measured curve with the simulated focus
**Sample** — `n` (linear index), `L` (thickness), `T_lin` (linear transmittance) | Parameter | Meaning | Units | Log units |
|-----------|----------------------------------------|----------|------------|
| `n_2` | 3rd-order nonlinear refractive index | m²/W | cm²/GW |
| `n_3` | 5th-order nonlinear refractive index | m⁴/W² | cm⁴/GW² |
| `n_4` | 7th-order nonlinear refractive index | m⁶/W³ | cm⁶/GW³ |
| `alpha_2` | Two-photon absorption coefficient | m/W | cm/GW |
| `alpha_3` | Three-photon absorption coefficient | m³/W² | cm³/GW² |
| `alpha_4` | Four-photon absorption coefficient | m⁵/W³ | cm⁵/GW³ |
**Laser** — `E` (pulse energy), `lambda_`, `w_0` (waist), `tau_FWHM`, `M2` ### z-position offsets
**Measurement** — `S` (aperture transmittance), `d` (focus-to-aperture distance) `Delta_z_OA`, `Delta_z_CA`, `Delta_z_OACA` (mm) shift the OA, CA, and CA/OA
z-axes independently so that measured and simulated curves line up when the two
detector arms are not perfectly co-registered.
**Error estimates** — `E_err`, `w_err`, `tau_err`, as fractions, propagated into ### Sample parameters
the reported uncertainty
**Simulation grid** — `z_points`, `r_points`, `r_max`, `t_points`, `t_max`, | Parameter | Meaning | Units |
`ra_points`, `endpts_percent`, `max_iterations` |-------------|---------------------------------------------|-------|
| `n` | Linear refractive index | — |
| `L` | Sample thickness | m |
| `T_lin` | Linear transmittance | — |
</details> ### Laser parameters
| Parameter | Meaning | Units |
|-------------|---------------------------------------------|-------|
| `E` | Pulse energy | J |
| `lambda` | Wavelength | m |
| `w_0` | Beam waist, HW1/e² of irradiance | m |
| `tau_FWHM` | Pulse duration, FWHM of irradiance | s |
| `M2` | Beam quality parameter M² | — |
### Measurement parameters
| Parameter | Meaning | Units |
|-------------|---------------------------------------------|-------|
| `S` | Linear transmission of the aperture | — |
| `d` | Distance from focus to aperture | m |
### Error estimates
`E_err`, `w_err`, and `tau_err` are fractional uncertainties on the energy,
waist, and pulse duration; they are combined in quadrature into `Tot_err` and
written to the log as the systematic uncertainty on the extracted coefficients.
— `E_err`, `w_err`, `tau_err`, as fractions, propagated into the reported
uncertainty
### Simulation parameters
| Parameter | Default | Meaning |
|-------------|---------|-----------------------------------------------------|
| `z_points` | `75` | Number of z-scan points (must be even) |
| `r_points` | `500` | Radial points at the sample |
| `r_max` | `3` | Radial integration limit, in beam radii *w* |
| `t_points` | `50` | Temporal points |
| `t_max` | `3` | Temporal integration limit, in pulse widths *τ* |
| `ra_points` | `50` | Radial points at the closed aperture |
| `endpts_%` | `10` | Percentage of edge points to use for normalization |
| `max_iter` | `100` | Maximum iterations for the parameter refinement |
Memory use scales as `r_points x ra_points x t_points x z_points`, so raising Memory use scales as `r_points x ra_points x t_points x z_points`, so raising
the grid resolution raises RAM sharply; the interface shows the estimate as you the grid resolution raises RAM sharply; the interface shows the estimate as you