Z-Scan Beam Simulation

License: GPLv3

A desktop application for fitting Z-scan measurements of nonlinear optical materials. It loads raw open-aperture (OA) and closed-aperture (CA) scan data, applies reference-channel correction and baseline normalization, simulates the expected transmittance from a set of nonlinear coefficients, and fits those coefficients to the measurement.

The Z-scan technique measures nonlinear refraction and absorption by translating a thin sample through the focus of a pulsed laser beam and recording transmitted energy as a function of position. This program models the beam propagating through the sample and on to a far-field aperture, and refines the nonlinear coefficients until the simulated curves match the measured ones.

The method follows Ferdinandus et al., Measuring small solute nonlinearities in solution by dual-arm Z-Scan technique (FiO 2012, paper FM4I.6) — see Citing this work.

Fit results with measured data

Features

  • Two operation modes. Fitting imports a measured scan and compares the simulation with it. Simulation calculates and displays the signals with no data at all, over a z range you specify.
  • Load raw scan files with configurable 1-indexed column mapping, so data from different acquisition setups can be read without editing the file. A header row is skipped. Leaving the delimiter blank splits on whitespace, which covers tab- and space-separated exports; set it explicitly (,) for CSV.
  • Reference-channel correction to divide out shot-to-shot laser fluctuation, plus an optional frequency channel.
  • Baseline normalization by a linear fit through the wings of the scan, where the sample is far from focus and the response should be flat. The fraction of points taken from each end is adjustable.
  • Nonlinear refraction and absorption to third order: n₂, n₃, n₄ and α₂, α₃, α₄.
  • Full beam simulation: Gaussian beam with M² correction, temporal pulse integration, radial propagation to the aperture, and both open- and closed-aperture transmittance.
  • Automatic fitting — tick any subset of the six nonlinear coefficients and refine them by Nelder-Mead simplex against the measured curves.
  • Fit quality reporting: chi-square per curve, peak-to-valley comparisons between data and fit, and a propagated uncertainty estimate from the energy, spot-size and pulse-width errors.
  • Export the fit, a full parameter log, and publication-ready plots.
  • Settings persistence — every field is saved on exit and restored on the next launch.

Screenshots

The left panel holds every processing and physical parameter; results appear in three tabs — the referenced/normalized input data, the fit, and a log of all parameters and fit-quality numbers.

Referenced / normalized data. Raw and reference-corrected traces with the baseline fit drawn over them. The correction visibly removes laser fluctuation from both channels.

Reference plot

Log. Every parameter that produced the fit, plus the fit-quality numbers.

Log tab

Installation

Requires Python 3.8+ and:

numpy
scipy
matplotlib

Tkinter ships with most Python installs. On some Linux distributions it is a separate package:

sudo apt install python3-tk        # Debian / Ubuntu
sudo dnf install python3-tkinter   # Fedora

Then:

pip install numpy scipy matplotlib
python zscan_gui.py

Usage

  1. Choose the Operation Mode — Fitting (the default) or Simulation. The steps below describe Fitting; for Simulation see Simulation mode.
  2. Browse to a scan file and set the column numbers for position, CA, OA and the reference channel. Columns are 1-indexed, matching how they are usually numbered in acquisition software.
  3. Import. The Referenced / Normalized tab shows the raw and corrected traces with the baseline fit drawn over them, so you can confirm the columns and flags are right before simulating. Changing a column or a processing flag re-reads the file and redraws immediately.
  4. Enter the measurement parameters — pulse energy, wavelength, beam waist, pulse width, sample thickness and refractive index, aperture transmittance.
  5. Run Simulation to overlay the simulated curves on the data.
  6. Refine Fit — tick the coefficients to vary and let the optimizer adjust them. It runs in the background with a live iteration count.
  7. Export to write the fit data, log and images next to the input file.

Simulation mode

Selecting Simulation calculates the signals without any measurement. Every input that describes a data file is switched off — the Data File, Data Columns and Processing Flags boxes, along with Import, Refine Fit, Max iterations and the Referenced / Normalized tab — and Run Simulation becomes available with nothing imported.

The scan's extent is not measured but stated: z range (sim.) sets the half-width in Rayleigh ranges z₀ either side of focus (default 15). The Fit Results panels show the simulated curves alone, padded by 5% above and below, and the log reports the simulated peak-to-valley values in place of the data-versus-fit comparison and chi-squares.

Show CA (else OA/CA) stays available, since it selects which curve the right-hand panel draws and both are simulated. Ticking it repeats the run so the plot follows immediately.

Anything already imported is kept while you work in Simulation, so switching back to Fitting does not mean importing again. Switching either way clears the results on screen, which belonged to the other mode.

Exported files

Written alongside the input file, with an optional extra label in the name. In Simulation mode there is no input file to name them from, so Export asks where to write; the CSV then holds the simulated curves and no reference figure is produced:

File Contents
<name>_fit.<ext> CSV of position, data and fit for the OA, CA and CA/OA curves; in Simulation, position and the simulated OA, CA and CA/OA
<name>_fit_log.txt Every parameter, fit-quality metric and output path
<name>_fit.jpg / .svg Fit figure, raster and vector
<name>_fit_ref.jpg / .svg Referenced/normalized figure (Fitting only)

Project structure

The program is split so the physics can be used without the interface:

  • zscan_core.py — the simulation engine. No GUI or pyplot state; figures are built as matplotlib.figure.Figure objects so any front end can embed them. Import it directly to drive the simulation from a script or notebook:

    from dataclasses import replace
    from zscan_core import load_settings, load_and_normalize_data, run_simulation
    
    cfg = replace(load_settings("last_settings.json"), import_filepath="scan.txt")
    DATA, norm_OA, norm_CA = load_and_normalize_data(cfg)
    results = run_simulation(cfg, DATA)
    print(results["OA_chi_square"], results["deltaOAcomp"])
    

    Every parameter lives in a single ZScanConfig dataclass, which serializes to and from JSON, so a fit is fully reproducible from its settings file.

  • zscan_gui.py — the Tkinter interface. Simulation and fitting run on background threads so the window stays responsive, and plots rescale with it.

Parameters

Operation mode

Option Effect
Fitting Default. Imports a scan and compares the simulation with it.
Simulation No data: the calculated signals are shown alone, over z range (sim.). The data-file inputs, Import, Refine Fit, Max iterations and the Referenced / Normalized tab are disabled.

Data columns (1-indexed)

Variable Default Column contents
z_col 0 Sample position z (mm)
CA_col 1 Closed-aperture detector signal (V)
FREQ_col 3 Frequency / shot-index channel
OA_col 4 Open-aperture detector signal (V)
REF_col 7 Reference detector signal (V)
FREQ_col 9 Lock-in frequency column (V)

Processing flags

Parameter Meaning
use_ref Apply the reference-channel correction
use_freq Appliy the frequency-channel correction
flip_z 1 reverses the sign of the z axis
do_norm_OA, do_norm_CA Fit and divide out the linear wing baseline
flip_OA, flip_CA, flip_REF Invert a raw channel about its median (0 or 1)
flip_OA_disp, flip_CA_disp Invert the normalized traces about 1 for display/fitting
show_CA Plot CA in the right panel; if False, plot CA/OA instead

Nonlinear coefficients

Parameter Meaning Units Log units
n_2 3rd-order nonlinear refractive index m²/W cm²/GW
n_3 5th-order nonlinear refractive index m⁴/W² cm⁴/GW²
n_4 7th-order nonlinear refractive index m⁶/W³ cm⁶/GW³
alpha_2 Two-photon absorption coefficient m/W cm/GW
alpha_3 Three-photon absorption coefficient m³/W² cm³/GW²
alpha_4 Four-photon absorption coefficient m⁵/W³ cm⁵/GW³

z-position offsets

Delta_z_OA, Delta_z_CA, Delta_z_OACA (mm) shift the OA, CA, and CA/OA z-axes independently so that measured and simulated curves line up when the two detector arms are not perfectly co-registered.

Sample parameters

Parameter Meaning Units
n Linear refractive index —
L Sample thickness m
T_lin Linear transmittance —

Laser parameters

Parameter Meaning Units
E Pulse energy J
lambda Wavelength m
w_0 Beam waist, HW1/e² of irradiance m
tau_FWHM Pulse duration, FWHM of irradiance s
M2 Beam quality parameter M² —

Measurement parameters

Parameter Meaning Units
S Linear transmission of the aperture —
d Distance from focus to aperture m

Error estimates

E_err, w_err, and tau_err are fractional uncertainties on the energy, waist, and pulse duration; they are combined in quadrature into Tot_err and written to the log as the systematic uncertainty on the extracted coefficients. — E_err, w_err, tau_err, as fractions, propagated into the reported uncertainty

Simulation parameters

Parameter Default Meaning
z_points 76 Number of z-scan points (must be even)
r_points 500 Radial points at the sample
r_max 3 Radial integration limit, in beam radii w
t_points 50 Temporal points
t_max 3 Temporal integration limit, in pulse widths τ
ra_points 50 Radial points at the closed aperture
endpts_% 10 Percentage of edge points to use for normalization
z range (sim.) 15 Half-width of the simulated scan, in Rayleigh ranges z₀. Simulation mode only
max_iter 100 Maximum iterations for the parameter refinement (Fitting only)

Memory use scales as r_points x ra_points x t_points x z_points, so raising the grid resolution raises RAM sharply; the interface shows the estimate as you edit those fields.

Citing this work

If this program contributes to work you publish, please cite the paper the method comes from:

M. R. Ferdinandus, M. Reichert, T. R. Ensley, D. A. Fishman, S. Webster, D. J. Hagan, and E. W. Van Stryland, "Measuring small solute nonlinearities in solution by dual-arm Z-Scan technique," in Frontiers in Optics 2012/Laser Science XXVIII, OSA Technical Digest (online) (Optica Publishing Group, 2012), paper FM4I.6. https://doi.org/10.1364/FIO.2012.FM4I.6

@inproceedings{Ferdinandus2012DualArmZScan,
  author    = {Ferdinandus, Manuel R. and Reichert, Matthew and
               Ensley, Trenton R. and Fishman, Dmitry A. and Webster, Scott and
               Hagan, David J. and Van Stryland, Eric W.},
  title     = {Measuring small solute nonlinearities in solution by dual-arm
               {Z}-Scan technique},
  booktitle = {Frontiers in Optics 2012/Laser Science XXVIII},
  series    = {OSA Technical Digest (online)},
  publisher = {Optica Publishing Group},
  year      = {2012},
  pages     = {FM4I.6},
  doi       = {10.1364/FIO.2012.FM4I.6},
}

The code was written by the paper's first author. It was ported from a MATLAB implementation of this method, then restructured into a separate simulation engine and interface; the physics and array shapes follow the original derivation.

License

Copyright (C) 2019-2026 Manuel R. Ferdinandus

This program is free software: you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation, either version 3 of the License, or (at your option) any later version.

This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the GNU General Public License for more details.

You should have received a copy of the GNU General Public License along with this program. If not, see https://www.gnu.org/licenses/.

The full license text is in LICENSE — download it from https://www.gnu.org/licenses/gpl-3.0.txt and save it as LICENSE in the repository root. Both source files carry the corresponding notice at the top.

NumPy, SciPy and Matplotlib are all BSD-licensed, so distributing this program under the GPL alongside them poses no conflict.

S
Description
No description provided
Readme GPL-3.0
1.1 MiB
Languages
Python 100%