360 lines
18 KiB
Markdown
360 lines
18 KiB
Markdown
# Changelog
|
|
|
|
All notable changes to BD Fresnel are documented in this file.
|
|
|
|
The format is based on [Keep a Changelog](https://keepachangelog.com/en/1.1.0/).
|
|
|
|
## [3.0] - 2026-11-14
|
|
|
|
### Added
|
|
|
|
- **Semi-analytical** algorithm, selectable from a third toggle in the
|
|
Operation mode box alongside **Fresnel propagation**:
|
|
- Instead of propagating the exit field to the detector by FFT, it
|
|
uses closed-form expressions for the power difference and total at
|
|
the detector, already integrated over the detector plane. Only an
|
|
integral over time at each delay remains.
|
|
- The time window runs +/- 5T about each delay, where T = tau_p/tau_e.
|
|
- Transmission is the energy total normalised to its value before the
|
|
pump arrives: the first delay point when fitting, and
|
|
min(-3 rho, -3) tau_e when simulating, so a thin sample with
|
|
rho -> 0 still normalises away from the peak.
|
|
- `X_num` and `X_max` are disabled in this mode: the expressions are
|
|
already integrated over space, so there is no spatial grid.
|
|
- Applies to both measurement modes. In pump probe the probe is
|
|
centred, x_0 = 0, so the deflection term vanishes identically and
|
|
the transmission alone is calculated.
|
|
- Both algorithms are kept so results can be compared. Switching
|
|
discards the current result, which belongs to the other method.
|
|
- The algorithm is saved with the other settings and recorded in the
|
|
parameter log.
|
|
- The absorptive term in the deflection expression carries a
|
|
PROVISIONAL coefficient, `SA_ABSORPTIVE_FACTOR`. As derived, that
|
|
term matched neither the sign nor the magnitude of the Fresnel
|
|
result: a purely absorptive mechanism disagreed by a factor of
|
|
about -69, with correlation exactly -1. Dividing by
|
|
D = d / (pi w_0_p^2 / lambda_p) = d / z_0_p accounts for the distance and
|
|
wavelength dependence exactly, and is independent of sample
|
|
thickness at constant alpha_2_exp*l; the remaining -1/(1 + 2W^2) was
|
|
measured against the Fresnel path rather than derived. With it the
|
|
two methods agree to 0.5% for mechanisms with a refractive part and
|
|
4.5% for purely absorptive ones. It is isolated in one function
|
|
pending a corrected derivation.
|
|
|
|
- The parameter log's `Calculated Parameters (pump)` section reports
|
|
`sigma_e`, the normalised pump linear absorption `alpha_e * L / 2`.
|
|
It sits next to `T_lin`, which restates it: with no pump two-photon
|
|
absorption `T_lin = (1 - R_p) exp(-2 sigma_e)`. Read from the
|
|
parameters rather than the result, so it is reported for either
|
|
algorithm.
|
|
|
|
- The `Other` panel and the parameter log's `Other` section both report
|
|
`eta (peak)` and `Gamma (peak)`, the largest nonlinear phase and
|
|
absorption exponent anywhere on the grid, from whichever algorithm
|
|
produced the result. These are the only readouts in those panels that
|
|
come from the result rather than from the parameters, since they are
|
|
peaks over the whole (T, T_d) grid and cannot be had without running
|
|
the calculation. They therefore show `--` until a Fit or Simulate
|
|
completes, and return to `--` whenever that result is discarded, so a
|
|
stale phase is never displayed beside freshly edited parameters.
|
|
`eta (peak)` is the yardstick for the semi-analytical algorithm's
|
|
validity: its `dP` is first order in the response while Fresnel
|
|
propagation keeps every order, so the two agree only while the phase
|
|
is small. On the 4-layer mechanisms the peak of `DeltaEoE` sits at
|
|
the same delay to within 0.0003 tau_e at `eta (peak)` = 0.14, but
|
|
0.0035 at 1.41, 0.077 at 5.6 and 0.36 at 14.1. Holding `eta (peak)`
|
|
fixed, the agreement is unaffected by sample thickness (0.05 to
|
|
2 mm), by walk-off (`rho` 0.04 to 1.7) and by pump absorption
|
|
(`sigma_e` 0 to 10), so this one number sets the regime. Both
|
|
algorithms report it to within 2e-5 of each other.
|
|
|
|
### Changed
|
|
|
|
- **Both beams are now explicitly subscripted: the pump carries `_e`
|
|
(excitation) and the probe carries `_p`.** This restores the
|
|
convention the original MATLAB used, which the Python port had
|
|
partly inverted (`_p` for the pump) and partly dropped (no suffix
|
|
for the probe). No beam quantity is unsubscripted any more; the only
|
|
unsuffixed parameters left are the ones belonging to neither beam --
|
|
the sampling grid, the sample geometry (`l`, `d`, `shift`), the
|
|
relative offset `x_0`, and the walk-off `rho`.
|
|
|
|
| pump: was | now | | probe: was | now |
|
|
|-----|-----|-|-----|-----|
|
|
| `E_p` | `E_e` | | `tau_FWHM` | `tau_p_FWHM` |
|
|
| `tau_p_FWHM` | `tau_e_FWHM` | | `w_0` | `w_0_p` |
|
|
| `w_0p` | `w_0_e` | | `lambda_` | `lambda_p` |
|
|
| `lambda_p` | `lambda_e` | | `R` | `R_p` |
|
|
| `R_p` | `R_e` | | `alpha` | `alpha_p` |
|
|
| `alpha_p` | `alpha_e` | | `n` | `n_p` |
|
|
| `n_p` | `n_e` | | `dndl` | `dndl_p` |
|
|
| `dndl_p` | `dndl_e` | | `tau` | `tau_p` |
|
|
| `alpha_2d` | `alpha_2d_e` | | `ng` | `ng_p` |
|
|
| `tau_p` | `tau_e` | | `z_0` | `z_0_p` |
|
|
| `sigma_p` | `sigma_e` | | `Tlin` | `Tlin_p` |
|
|
| `I_0_p` | `I_0_e` | | | |
|
|
| `ng_p` | `ng_e` | | | |
|
|
| `z_0_p` | `z_0_e` | | | |
|
|
| `Tlin_p` | `Tlin_e` | | | |
|
|
| `sigma_p2` / `sigma_p20` | `sigma_e2` / `sigma_e20` | | | |
|
|
| `a_p_0`, `W_0_p`, `I_p_pulse` | `a_e_0`, `W_0_e`, `I_e_pulse` | | | |
|
|
|
|
Note that six probe names -- `tau_p_FWHM`, `lambda_p`, `R_p`,
|
|
`alpha_p`, `n_p`, `dndl_p` -- are spelled exactly like what were
|
|
**pump** names before this release. Reading an old log or settings
|
|
file alongside a new one needs care; see the settings note below for
|
|
how the program itself is protected.
|
|
|
|
The pump pass renamed 389 identifier occurrences textually; the
|
|
probe pass was done by rewriting 116 exact AST node positions
|
|
instead, because `n`, `R`, `alpha` and `tau` are overloaded in this
|
|
codebase (lambda parameters, the author's initials in the copyright
|
|
line, the semi-analytical formula's geometric `R`, a diagnostics
|
|
dict key) and no text pattern distinguishes them.
|
|
|
|
| was | now | | was | now |
|
|
|-----|-----|-|-----|-----|
|
|
| `E_p` | `E_e` | | `tau_p` | `tau_e` |
|
|
| `tau_p_FWHM` | `tau_e_FWHM` | | `sigma_p` | `sigma_e` |
|
|
| `w_0p` | `w_0_e` | | `I_0_p` | `I_0_e` |
|
|
| `lambda_p` | `lambda_e` | | `ng_p` | `ng_e` |
|
|
| `R_p` | `R_e` | | `z_0_p` | `z_0_e` |
|
|
| `alpha_p` | `alpha_e` | | `Tlin_p` | `Tlin_e` |
|
|
| `n_p` | `n_e` | | `sigma_p2` | `sigma_e2` |
|
|
| `dndl_p` | `dndl_e` | | `sigma_p20` | `sigma_e20` |
|
|
| `alpha_2d` | `alpha_2d_e` | | `a_p_0` | `a_e_0` |
|
|
|
|
`w_0p` gains the separating underscore it was missing, so the field
|
|
name finally matches the variable that held it (`w_0_p`) and its
|
|
siblings `z_0_e` and `I_0_e`. `alpha_2d_e` keeps the `d` for
|
|
degenerate.
|
|
|
|
- **The Electronic mechanism's parameters are renamed `alpha_2e` ->
|
|
`alpha_2el` and `n_2e` -> `n_2el`**, since their `e` meant
|
|
*electronic* and would otherwise collide with `_e` for excitation --
|
|
the pump's degenerate two-photon coefficient would have wanted the
|
|
name `alpha_2e`, which was taken. `_e` now means excitation and
|
|
nothing else.
|
|
|
|
- `BD_3_level_triplet`, `I_e_func` and `excite_prop` needed no
|
|
renaming at all: transliterated from the MATLAB, they already used
|
|
`_e` for the pump and `_p` for the probe, so `n_g_p`, `v_p` and
|
|
`I_p` inside them were correct all along. With both beams now
|
|
subscribed to that convention, the whole file reads against the
|
|
MATLAB without translation.
|
|
|
|
- **`last_settings.json` written by an earlier version is now refused
|
|
rather than misread.** Saved settings carry a `settings_schema`
|
|
stamp, and a file without the current one is rejected with a message
|
|
explaining why. This is not housekeeping: because six probe field
|
|
names are spelled exactly like former pump keys, loading an old file
|
|
by name alone would have installed the pump's wavelength, index,
|
|
reflectivity, absorption and dispersion as the *probe's* -- every
|
|
number plausible, every result wrong. Re-enter the parameters once
|
|
and they save in the new layout. Any other unrecognised key is still
|
|
dropped with a note on stderr.
|
|
|
|
- Log and panel labels follow their fields on both sides: `n_g_e`,
|
|
`sigma_e`, `T_lin_e`, `z_0_e`, `I_0_e` for the pump and `n_g_p`,
|
|
`T_lin_p`, `z_0_p` for the probe, with the beam parameters and
|
|
material blocks likewise. The GUI's pump calculated-parameters panel
|
|
also gains a `sigma_e` readout, which it had not displayed before.
|
|
|
|
- `T_lin_p` and `T_lin_e` are labelled `T_lin_p (transmission)` and
|
|
`T_lin_e (transmission)` in both the panel and the log, so the
|
|
quantity is named and not just abbreviated. The log's value column
|
|
moved from 25 to 29 to fit them, and the panel's label column from 20
|
|
to 23 characters -- which also squares up `z_0_p (Rayleigh range)`,
|
|
`z_0_e (Rayleigh range)` and `I_0_e (peak irradiance)`, all of which
|
|
had been overrunning the old 20 and pushing their own values out of
|
|
line.
|
|
|
|
- The rename is numerically inert: across 8 mechanism types x 2
|
|
algorithms x 3 pump-absorption configurations (49 cases, 3275
|
|
values), `DeltaEoE`, `T`, `Delta_n`, `eta_peak` and `Gamma_peak` are
|
|
bit-identical to the pre-rename code, and `pyflakes` reports zero
|
|
warnings before and after.
|
|
|
|
- Transmission y-axis tick labels are limited to three significant
|
|
digits, so "99.75" reads "99.8". The count is a floor rather than a
|
|
cap: a shallow scan spanning a few hundredths of a percent would
|
|
round every tick to "100", so digits are added only where that would
|
|
happen. The deflection axis is unaffected.
|
|
|
|
- The reference list in the README and the parameter log follows Optica
|
|
style: every author listed, initials first, abbreviated journal,
|
|
volume, full page range, year. Author lists and page ranges were
|
|
checked against the published papers.
|
|
- The duplicate supplementary-material citation is dropped, leaving two
|
|
references.
|
|
- The parameter log is written as UTF-8. The citations use typographic
|
|
quotes and an en dash, which the locale's encoding cannot always
|
|
represent.
|
|
|
|
### Fixed
|
|
|
|
- The instantaneous (Electronic) response for a sample with both
|
|
walk-off and pump absorption is evaluated through the scaled
|
|
complementary error function `erfcx` instead of the form in
|
|
`BD_Fresnel_func_ni.m`. That form multiplied `exp(sigma_e^2/rho^2 +
|
|
2 sigma_e u/rho)` by a difference of error functions falling to zero,
|
|
so the product was lost to cancellation well before it overflowed:
|
|
at `rho = 0.4336` (l = 0.5 mm) it was already wrong over 8.6% of the
|
|
delay axis at `sigma_e = 2.5` (alpha_e = 100 cm^-1), **returning
|
|
exactly 0.0 with no NaN to signal it**, and returned NaN above
|
|
`sigma_e = 3`. The usable ceiling on `alpha_e` for an Electronic
|
|
mechanism at that thickness was therefore about 60 cm^-1, not the
|
|
~120 where NaN first appeared. The rearrangement is algebraic, with
|
|
every surviving exponent non-positive; it reproduces the original
|
|
evaluated at 600 decimal digits to 1e-16..3e-13 relative over
|
|
`sigma_e` = 0.25 to 10 and `u` = -3 to 18, both where the original
|
|
works and where it does not.
|
|
- Pump absorption is applied to non-instantaneous mechanisms when the
|
|
walk-off is exactly zero. Substituting `T1 = (T+T_d) - rho +
|
|
rho*zeta` shows the `rho != 0` branch to be a depth average over
|
|
`atten_fac`, whose `rho -> 0` limit is the response at one instant
|
|
times the *mean* of `atten_fac`; the original used the bare response.
|
|
The dropped factor is `(1 - exp(-2 sigma_e))/(2 sigma_e)` for linear
|
|
pump absorption, so the response was previously **independent of
|
|
`alpha_e` altogether** in this case -- identical at `alpha_e` = 0 and
|
|
200 cm^-1. It was 1.27x too large at `alpha_e` = 10 cm^-1 and
|
|
l = 0.5 mm, and 10x at 200 cm^-1. The two branches now agree to nine
|
|
digits across `rho = 0`. This affects only samples with `ng_p` exactly
|
|
equal to `ng_e` (for example no dispersion entered at either
|
|
wavelength) *and* non-zero pump absorption; with `alpha_e = alpha_2d_e
|
|
= 0` the mean is exactly 1 and results are bit-identical to before.
|
|
- Pump absorption through the sample depth, `atten_fac`, is computed
|
|
from the closed-form solution of
|
|
`dI_e/dz = -alpha_e I_e - alpha_2d_e I_e^2` rather than the two-branch
|
|
expression carried over from `BD_Fresnel_func_ni.m`. That expression's
|
|
pump two-photon branch, `sqrt((B1 B2)/sigma_e)`, failed three ways: it
|
|
did not reduce to its own `sigma_e2 == 0` branch as `alpha_2d_e -> 0`
|
|
(it decayed at half the correct rate), it divided by zero whenever
|
|
`alpha_e` was zero, and it took the square root of a negative number
|
|
whenever `sigma_e20 > sigma_e`. The net effect was that **every run
|
|
combining a non-zero `alpha_2d_e` with a non-instantaneous mechanism
|
|
returned NaN at every delay** -- pump two-photon absorption worked
|
|
only for the Electronic mechanism, which does not use `atten_fac`.
|
|
Dropping the stray square root recovers that expression as the
|
|
linearisation of the exact one in `sigma_e20`, which is how the
|
|
intent was identified. Results for `alpha_2d_e = 0` are bit-identical
|
|
to before; the new form agrees with numerical integration of the
|
|
pump propagation equation to 2e-12 relative over `sigma_e` and
|
|
`sigma_e20` from 0 to 20, including both limits.
|
|
- A negative `alpha_2d_e` large enough to make the pump diverge inside
|
|
the sample (two-photon gain, no finite depth average) is now refused
|
|
with a message naming the largest `|alpha_2d_e|` that works for the
|
|
current thickness and pump irradiance, rather than silently returning
|
|
a meaningless attenuation of order 1e12.
|
|
- `mpmath` is imported only where it is used, in the pump two-photon
|
|
absorption term. Importing it at module level made an optional
|
|
dependency fatal at startup, so the program would not run at all
|
|
without it. If that term is ever reached without mpmath installed,
|
|
the error now says which package is missing and why.
|
|
|
|
## [2.6] - 2026-09-30
|
|
|
|
### Added
|
|
|
|
- **Simulation** run mode: a second toggle in the Operation mode box,
|
|
offering **Fitting** and **Simulation**. Fitting is the default and
|
|
behaves as before.
|
|
- Simulation mode, for computing a signal without a measurement:
|
|
- No file is imported. **Import...** and the data column and
|
|
delimiter fields are disabled.
|
|
- The delay range is entered as `T_d_min` and `T_d_max` (in `tau_p`)
|
|
in the Simulation parameters box. In Fitting mode these are taken
|
|
from the imported scan and the fields are disabled.
|
|
- Only the calculated curves are plotted, without data markers, and
|
|
the plot legend reads **Simulated** rather than Fit.
|
|
- The **Fit** button is relabelled **Simulate**, and is enabled by
|
|
valid parameters alone rather than by an imported file.
|
|
- The export holds the calculated signals on their own delay grid,
|
|
with no data columns to pair them with. The parameter log omits the
|
|
import path and the data column section.
|
|
- The mode combines with the measurement mode: Pump probe and
|
|
Simulation together give a single Transmission panel and a
|
|
two-column table, as Pump probe does when fitting.
|
|
- The run mode is saved with the other settings and recorded in the
|
|
parameter log.
|
|
|
|
### Changed
|
|
|
|
- The **Mode** box is renamed **Operation mode** and moved to the top of
|
|
the parameter panel, above Data columns. The parameter log and status
|
|
messages use the new name, and the log records both modes on one line.
|
|
- Changing run mode discards the current result, which described the
|
|
other mode's delay range or a measurement the new mode does not show.
|
|
Switching to Simulation sets the imported scan aside and switching
|
|
back reloads it. The run mode cannot be changed while a fit runs.
|
|
- `inputs_valid` takes a `require_data` argument, so Simulation can
|
|
enable its action button without an imported file. Fitting is
|
|
unaffected.
|
|
|
|
## [2.5] - 2026-09-21
|
|
|
|
### Added
|
|
|
|
- **Mode** box, below Data columns: a two-button toggle selecting
|
|
**Beam deflection** or **Pump probe**. Beam deflection is the default
|
|
and behaves as before.
|
|
- Pump probe mode, for measurements with the probe centred on the pump:
|
|
- Only the delay and transmission columns are required, so a
|
|
two-column scan can be imported. The DeltaEoE column is ignored and
|
|
its field disabled.
|
|
- `x_0` is fixed at 0 and cannot be edited.
|
|
- Only the Transmission plot is displayed and exported. The exported
|
|
image is named `<name>_fit_transmission.jpg` / `.svg` and is 3.25 in
|
|
wide, a single column. The exported data table holds delay, T (data)
|
|
and T (fit).
|
|
- The mechanism pull-downs offer the absorptive (NLA) and mixed
|
|
(both) mechanisms; a mixed mechanism contributes its absorptive
|
|
part. The purely refractive (NLR) mechanisms, which cannot change
|
|
the transmission, are excluded. Switching to pump probe empties any
|
|
slot holding one; switching back restores it, provided the slot is
|
|
still empty.
|
|
- The selected mode is saved with the other settings and recorded in the
|
|
parameter log.
|
|
|
|
### Changed
|
|
|
|
- Changing mode discards the current fit, which was computed with the
|
|
other mode's settings, and re-reads the imported file under the new
|
|
mode's column requirements. A file the new mode cannot use is cleared
|
|
with an explanation, and loaded again on switching back. The mode
|
|
cannot be changed while a fit runs.
|
|
- The source files carry the GPLv3 notice.
|
|
|
|
### Fixed
|
|
|
|
- Values such as `nan`, `inf` or `1e999` are rejected in parameter fields.
|
|
They previously turned the fit to NaN, and crashed the mechanism
|
|
parameters dialog.
|
|
- An invalid entry in a numeric field reverts to its previous value, as the
|
|
integer fields and the mechanism dialog already did, instead of becoming 0.
|
|
- Every parameter change is saved as it is made. Numeric fields, mechanism
|
|
choices and the diagnostics option were previously saved only on Import,
|
|
Fit, Export or Exit, so a crash lost them.
|
|
- Changing a column number or the delimiter re-reads only a file imported in
|
|
the current session. Straight after launch it could load the previous
|
|
session's file, or report an error about one that had since moved.
|
|
- When a file cannot be read with a new column number or delimiter, its data
|
|
and fit are cleared. They previously stayed on screen, and could be
|
|
exported, under settings that no longer described them.
|
|
- The data columns in use must each be different. Two quantities could
|
|
previously be read from the same column without any warning.
|
|
- The x tick label at the join between the Transmission and Deflection plots
|
|
is hidden when it collides with its neighbour. A rounding error let some
|
|
such labels through.
|
|
|
|
## [2.4] - 2026-09-21
|
|
|
|
First release recorded in this changelog. Earlier history is not documented
|
|
here.
|
|
|
|
### Fixed
|
|
|
|
- README parameter tables corrected to match the program's window: the
|
|
default T and DeltaEoE data columns were swapped, and several defaults,
|
|
labels and descriptions disagreed with what the interface shows.
|